Spectrophotometer YR07082

The UV-Visible measurement principle allows for the analysis of substances that absorb light at specific wavelengths. The procedure typically includes the preparation of a sample and a blank, the selection of reading conditions, and the recording of absorbance or transmittance. The specific characteristics of the YR07082 — such as wavelength range, bandwidth, optical system, accuracy, display, software, and included accessories — must be confirmed in the official technical data sheet before publication, as the specification text provided is empty.
This page provides an overview of the Advanced UV-Visible Spectrophotometer YR07082, including selection criteria, investment factors, operational advantages and maintenance recommendations. The aim is to support an informed decision before requesting a quotation.
This Kalstein model is available for laboratories across Latin America, Europe, Africa, the Middle East and Asia, with selection criteria adapted to the diversity of local purchasing, installation and support conditions in each region.
Key factors to consider

Before selecting the Advanced UV-Visible Spectrophotometer YR07082, confirm workflow, available space, required accessories and installation conditions. This prevents oversizing the purchase and helps prepare a more accurate quotation.
Its use is oriented to laboratory routines that require observation, control and operational continuity.
Across these markets, evaluation should consider varied local contexts such as international laboratory procurement spanning Latin America, Europe, Africa, the Middle East and Asia and match the product to the real workload of the laboratory in each country.
Market price range

The price of a UV-Visible spectrophotometer depends on the optical configuration, spectral range, accuracy, software, accessories, and included support. As a general reference, basic models can be found from approximately US$600–2,500, while professional bench-top equipment usually ranges around US$3,500–10,000. Advanced or high-end configurations can exceed US$13,000–20,000. These values are market references and do not represent the confirmed price of the YR07082; please request a quote to know the price, availability, and applicable delivery conditions.
The final price may vary according to configuration, accessories, logistics, warranty and territorial availability. Compare the technical sheet with the real laboratory workload before confirming the investment.
Budget comparison varies by country and should include the regulatory requirements specific to each market, the electrical conditions of each region, accessories, logistics and after-sales support.
Value for money

Equipment value depends on operational stability, required accuracy, ease of maintenance and accessory availability.
Value for money should be reviewed together with equipment stability, after-sales support, technical documentation and expected service life. A well-matched selection reduces downtime and improves return on investment.
Advantages and disadvantages

Its use is oriented to laboratory routines that require observation, control and operational continuity.
Main advantages include specialized application, integration with laboratory routines and accessory configuration. Limitations may include installation requirements, user training and preventive maintenance to preserve performance.
Expected lifespan based on use and maintenance

ModelYR07082Optical SystemSingle BeamLight SourceTungsten Lamp (W) and Deuterium Lamp (D₂)Wavelength Scan Range190–1100 nmSpectral Bandwidth1.8 nmDisplay7-inch TouchscreenWavelength Accuracy±0.5 nm at 656.1 nm (D₂)Wavelength Repeatability≤0.2 nmPhotometric Accuracy±0.3 %T (0–100 %T); ±0.002 Abs (0–0.5 Abs); ±0.004 Abs (0.5–1.0 Abs)Photometric Repeatability≤0.1 %T (0–100 %T); ≤0.001 Abs (0–0.5 Abs); ≤0.002 Abs (0.5–1.0 Abs)Reference Stability≤0.001 A/h at 250 nm and 500 nm, after 2 hours of warm-upPhotometric Range−0.3 to 3 A; 0–200 %T; 0–9999 C*Scattered Light≤0.05 %T at 220 nm and 360 nmOperating ModeStandalone or with optional PC softwareData OutputUSB; optional BluetoothPower SupplyAC 110/220 V, 50/60 HzEquipment Dimensions (L × W × H)47.
Service life depends on usage frequency, cleaning, storage, calibration when applicable and adherence to technical recommendations. A structured maintenance plan helps preserve accuracy, safety and operational continuity.
A support and spare-parts network deployed across Latin America, Europe, Africa, the Middle East and Asia helps preserve service continuity and plan preventive maintenance without extended downtime.
Specifications table for the Spectrophotometer YR07082

Spectrophotometer YR07082
The UV-Visible measurement principle allows for the analysis of substances that absorb light at specific wavelengths. The procedure typically includes the preparation of a sample and a blank, the selection of reading conditions, and the recording of absorbance or transmittance. The specific characteristics of the YR07082 — such as wavelength range, bandwidth, optical system, accuracy, display, software, and included accessories — must be confirmed in the official technical data sheet before publication, as the specification text provided is empty.
Market Price
The price of a UV-Visible spectrophotometer depends on the optical configuration, spectral range, accuracy, software, accessories, and included support. As a general reference, basic models can be found from approximately US$600–2,500, while professional bench-top equipment usually ranges around US$3,500–10,000. Advanced or high-end configurations can exceed US$13,000–20,000. These values are market references and do not represent the confirmed price of the YR07082; please request a quote to know the price, availability, and applicable delivery conditions.
Frequently Asked Questions
What does a UV-Visible spectrophotometer measure?
It measures the absorbance or transmittance of a sample at specific wavelengths, allowing for qualitative or quantitative analyses depending on the method used.
What samples can be analyzed?
It depends on the method, sample preparation, and available accessories. Generally, it can be used with liquid solutions compatible with UV-Visible analysis.
Can it be used outside the laboratory?
It should not be assumed that the YR07082 is portable. For fieldwork, confirm if the manufacturer allows use outside a controlled environment and review the requirements for power, temperature, protection, and transport. A common option is to take samples in the field and analyze them in a suitable laboratory or mobile lab.
What information do I need to request a quote?
Indicate the model YR07082, intended applications, type of samples, required accessories, and destination country. You can request the quote from the Kalstein team or generate an automatic quote from the Kalstein Plus platform.
Advantages and Considerations
Among the general advantages of UV-Visible spectrophotometry are the speed of reading, the ability to quantify substances using validated methods, and its usefulness in a wide variety of laboratory analyses. It can also facilitate the monitoring of processes and routine controls when working with consistent protocols.
As considerations, the result depends on sample preparation, method selection, and cuvette cleanliness. Turbid samples or those with particles can affect the reading, and some applications in the UV region require quartz cuvettes. Additionally, the performance and functions of the YR07082 should be evaluated against its official specifications before comparing it with other equipment or confirming its suitability for a specific method.
Product Use in the Field and Laboratory
In environmental, industrial, or process control scenarios, samples can be collected at the workplace and transported following the appropriate protocol for analysis. In mobile laboratories, the use of the equipment will depend on whether the conditions for power, temperature, vibration, and protection are compatible with the manufacturer's recommendations. For routine analysis, it is recommended to place it on a stable and level surface, work with defined methods, and record the measurement conditions along with the results.
Usage and Maintenance Recommendations
Before using the equipment, review the manual and confirm that the method, cuvettes, and accessories are compatible with the sample. Properly prepare the blank and samples, keep optical surfaces clean, and avoid handling cuvettes by the transparent faces. Schedule periodic performance checks according to the laboratory protocol and protect the instrument from dust, moisture, vibrations, and electrical variations. Confirm with Kalstein the installation, training, calibration, and maintenance requirements applicable to the YR07082.
| Model | YR07082 |
|---|---|
| Optical System | Single Beam |
| Light Source | Tungsten Lamp (W) and Deuterium Lamp (D₂) |
| Wavelength Scan Range | 190–1100 nm |
| Spectral Bandwidth | 1.8 nm |
| Display | 7-inch Touchscreen |
| Wavelength Accuracy | ±0.5 nm at 656.1 nm (D₂) |
| Wavelength Repeatability | ≤0.2 nm |
| Photometric Accuracy | ±0.3 %T (0–100 %T); ±0.002 Abs (0–0.5 Abs); ±0.004 Abs (0.5–1.0 Abs) |
| Photometric Repeatability | ≤0.1 %T (0–100 %T); ≤0.001 Abs (0–0.5 Abs); ≤0.002 Abs (0.5–1.0 Abs) |
| Reference Stability | ≤0.001 A/h at 250 nm and 500 nm, after 2 hours of warm-up |
| Photometric Range | −0.3 to 3 A; 0–200 %T; 0–9999 C* |
| Scattered Light | ≤0.05 %T at 220 nm and 360 nm |
| Operating Mode | Standalone or with optional PC software |
| Data Output | USB; optional Bluetooth |
| Power Supply | AC 110/220 V, 50/60 Hz |
| Equipment Dimensions (L × W × H) | 47.5 × 37 × 23 cm |
| Net Weight | 17 kg |
| Packaging Dimensions | 62 × 53 × 37 cm |
| Gross Weight of Packaging | 20 kg |
| Model | YR07082 |
|---|---|
| Optical System | Single Beam |
| Light Source | Tungsten Lamp (W) and Deuterium Lamp (D₂) |
| Wavelength Scan Range | 190–1100 nm |
| Spectral Bandwidth | 1.8 nm |
| Display | 7-inch Touchscreen |
| Wavelength Accuracy | ±0.5 nm at 656.1 nm (D₂) |
| Wavelength Repeatability | ≤0.2 nm |
| Photometric Accuracy | ±0.3 %T (0–100 %T); ±0.002 Abs (0–0.5 Abs); ±0.004 Abs (0.5–1.0 Abs) |
| Photometric Repeatability | ≤0.1 %T (0–100 %T); ≤0.001 Abs (0–0.5 Abs); ≤0.002 Abs (0.5–1.0 Abs) |
| Reference Stability | ≤0.001 A/h at 250 nm and 500 nm, after 2 hours of warm-up |
| Photometric Range | −0.3 to 3 A; 0–200 %T; 0–9999 C* |
| Scattered Light | ≤0.05 %T at 220 nm and 360 nm |
| Operating Mode | Standalone or with optional PC software |
| Data Output | USB; optional Bluetooth |
| Power Supply | AC 110/220 V, 50/60 Hz |
| Equipment Dimensions (L × W × H) | 47.5 × 37 × 23 cm |
| Net Weight | 17 kg |
| Packaging Dimensions | 62 × 53 × 37 cm |
| Gross Weight of Packaging | 20 kg |
Products related to model YR07082

Ultramicro Spectrophotometer YR07037 – YR07038
Ver producto
Ultramicro Spectrophotometer YR07039
Ver producto
Matrix Spectrophotometer YR07018
Ver producto
YR06702//YR06702-3 Series Single and Dual Beam UV-Visible Spectrophotometer
Ver producto
Array Spectrophotometer YR06664
Ver producto
Non-contact Benchtop Spectrophotometer YR05480
Ver producto
Biophotometer Micro Spectrophotometer YR06033
Ver producto
Spectrophotometer double beam YR01862-1 - YR01862-2
Ver productoFrequently Asked Questions about the Spectrophotometer YR07082
What does a UV-Visible spectrophotometer measure?
It measures the absorbance or transmittance of a sample at specific wavelengths, allowing for qualitative or quantitative analyses depending on the method used.
What samples can be analyzed?
It depends on the method, sample preparation, and available accessories. Generally, it can be used with liquid solutions compatible with UV-Visible analysis.
Can it be used outside the laboratory?
It should not be assumed that the YR07082 is portable. For fieldwork, confirm if the manufacturer allows use outside a controlled environment and review the requirements for power, temperature, protection, and transport. A common option is to take samples in the field and analyze them in a suitable laboratory or mobile lab.
What information do I need to request a quote?
Indicate the model YR07082, intended applications, type of samples, required accessories, and destination country. You can request the quote from the Kalstein team or generate an automatic quote from the Kalstein Plus platform.
How do I choose the right Spectrophotometer YR07082 for my laboratory?
Selection should start from workload, critical technical parameters, available space and frequency of use. Kalstein can support the comparison between operational needs, budget and maintenance requirements to avoid choosing equipment that is oversized or insufficient.
What should I review before requesting a quotation?
Review the model, technical specifications, required accessories, electrical conditions, consumables, delivery timing and installation needs. With this information, the quotation can be prepared more accurately and aligned with the laboratory workflow.
Does Kalstein provide guidance for this type of equipment?
Yes. The commercial and technical team can support product evaluation, explain differences between models, answer application questions and advise on preventive maintenance, spare parts and documentation required for a secure purchase.
What maintenance helps extend the equipment lifespan?
Follow the manufacturer manual, perform regular cleaning, verify calibrations or controls when applicable, avoid workload abuse and keep maintenance records. Stable use and preventive routines reduce failures and improve operational continuity.