Product Description
The YR06214 Infant Incubator is a state-of-the-art piece of medical equipment designed to provide the ideal environment for infant care. It is specifically engineered to offer accurate and reliable thermal management, ensuring newborns remain in optimal conditions for their health and development. The incubator operates on an electrical requirement of ~220V 50Hz and utilizes heater power of 550VA, making it both energy-efficient and powerful enough to maintain consistent temperatures.
This incubator features distinct air and skin mode temperature ranges, with air mode extending from 25.0°C to 37.0°C and skin mode from 34.0°C to 37.0°C, addressing the specific needs of different infant conditions. The YR06214 also includes precise temperature alarms to ensure safety; the air mode temperature alarm triggers at temperatures below 38.0°C, while the skin mode alarm is activated if the temperature exceeds 40.0°C, safeguarding against potentially harmful temperature fluctuations.
| Model | YR06214 |
|---|---|
| Electrical requirement | ~220V 50Hz |
| Heater Power | 550VA |
| Air mode temperature range | 25.0°C-37.0°C |
| Skin mode temperature range | 34.0°C-37.0°C |
| Air mode temperature alarm | <38.0°C |
| Skin mode temperature alarm | < 40.0°C (Air temperature) |
| Air mode high temperature range | 37.1°C~39.0°C |
| Skin mode high temperature range | 37.1°C~38.0°C |
| Hight temperature mode alarm | < 40.0°C (Air temperature) |
| Air mode temperature display range | 20.0°C-45.0°C |
| Skin mode temperature display range | 25.0°C~45.0°C |
| Air mode alarm | ±3.0°C |
| Skin mode alarm | ±1.0°C |
| Skin temperature sensor accuracy | ±0.3°C |
| Temperature control accuracy | ≤0.5°C |
| Mattress temperature uniformity | ≤0.8°C |
| Humidity control range | 30-90%RH |
| Humidity display range | 0-99.9%RH |
Market Price
When it comes to the cost of the YR06214 Infant Incubator, the market price generally falls within a range of $12,500 to $13,300 USD. Investing in this essential piece of neonatal care equipment aligns with the top-tier functionality and features it offers, balancing price with long-term value. Hospitals and clinics seeking reliable and advanced incubators will find this price range reflective of the incubator’s capabilities.
This financial outlay ensures healthcare providers can offer the highest standard of care in neonatal wards. With costs varying slightly based on suppliers and any additional features or service agreements, it is recommended to secure a personalized quote for the best assessment of your needs and budget. To facilitate this process, consider utilizing the Kalstein Plus platform for an instant quotation to explore potential packages and discounts.
Frequently Asked Questions
What is the power requirement for the YR06214 Incubator?
The YR06214 Incubator runs on a ~220V 50Hz electrical supply and requires 550VA for heater power. This ensures efficient and consistent operation across variable conditions.
How does the temperature alarm system work?
The incubator is equipped with robust alarm systems for both air and skin modes. It alerts caregivers if temperatures fall below 38.0°C in air mode or exceed 40.0°C linked with the skin mode. This safety feature is crucial to maintain a stable and conducive environment for neonates.
Advantages and Disadvantages
Advantages:
The YR06214 Incubator offers precise temperature control, which is crucial for the well-being of premature or ill neonates. It also provides reliable humidity control, ensuring the infant’s environment is optimal for development. The advanced alarm system enhances safety by quickly alerting caregivers to any abnormalities.
Disadvantages:
While the YR06214 is packed with advanced features, it may require professional installation and setup to ensure all specifications are met appropriately. Additionally, due to the high-tech nature of the incubator, specialized training for healthcare staff might be necessary to fully leverage its capabilities.
Use of the Product in the Field
In practical settings, the YR06214 Infant Incubator is utilized across various hospital departments, prominently in neonatal intensive care units (NICUs). It plays a pivotal role in creating a controlled microenvironment, closely replicating the uterine conditions required for the growth and recuperation of premature infants.
The incubator is designed to offer consistent monitoring and adjustment of temperature and humidity levels tailored to the needs of each infant, facilitating ease of access for medical staff during interventions without exposing infants to external environmental changes.
Recommendations
For optimal performance, it’s integral to keep the YR06214 Incubator in a location with stable power supply conditions and ensure it undergoes regular maintenance checks. It’s recommended to have backup sensors and alarms to maintain functionality even in the unlikely event of device component failures.
Engage qualified personnel for both setup and periodic reviews, and ensure that all medical staff are adequately trained in its use. By doing so, healthcare providers can maximize the lifespan and functionality of this sophisticated incubator, ensuring the highest standard of neonatal care.
If you’re looking for a blend of innovation and quality, you’ve come to the right place. At Kalstein, we offer you the luxury of exploring our exclusive catalog of laboratory equipment. We manufacture each piece with a level of excellence. Our intuitive and agile online purchasing channels are designed for your convenience, guaranteeing the most friendly prices. Don’t hesitate any longer; we bring science to life; it’s time to be part of our community. https://kalstein.net/en/product/infant-incubator-yr06214/.