Advanced Technical Excellence: Class II Biological Safety Cabinets

Product Description

The YR05704 to YR05707 Class II A2 Biological Safety Cabinets by Kalstein stand as epitomes of precision in laboratory safety, designed to offer an uncompromised shield against biological hazards. These cabinets are engineered to provide Class 100 cleanliness according to US Federal 209E standards, ensuring that sensitive operations are carried out in safe and sterile environments.

Featuring exceptional airflow management, each model offers consistent average wind speeds inside the door (0.38±0.025 m/s), at the middle (0.26±0.025 m/s), and inside (0.27±0.025 m/s). These velocities are critical in curbing contamination and maintaining the integrity of sensitive experiments. Furthermore, the front suction wind speed at 0.55m±0.025m/s (30% efflux) adds another layer of protection, ensuring contaminated air is effectively channeled out.

Model YR05704 YR05705 YR05706 YR05707
Cleanliness Class 100@≥0.5μm (US Federal 209E)
Number of colonies ≤0.5pcs/dish·hour (Φ90mm culture plate)
Average wind speed Inside the door 0.38±0.025m/s
Middle 0.26±0.025m/s
Inside 0.27±0.025m/s
Front suction wind speed 0.55m±0.025m/s(30% Efflux)
Noise ≤65dB(A)
Voltage AC single-phase 220V/50Hz
Maximum power consumption 400W 500W 600W 700W
Weight 170KG 210KG 250KG 270KG
Working area W1×D1×H1 840×650×620 1040×650×620 1340×650×620 1640×650×620
Overall dimension W×D×H 1000×800×2100 1200×800×2100 1500×800×2100 1800×800×2100
High efficiency filter specification and quantity 780×490×50× 1 520×380×70× 1 820×380×70× 1 1120×380×70× 1
Specification and quantity of fluorescent lamp/ultraviolet lamp 12W×1 / 20W×1 20W×2/20W×1 30W×2/30W×1 30W×2/30W×1

Market Price

The market price for Class II Biological Safety Cabinets can range significantly based on specifications and features. Kalstein offers a competitive price range, and you can expect these cabinets to be available between $5,500 and $6,600 USD. This competitive price point reflects the superior quality and advanced technologies integrated into our models, designed for those who prioritize safety and reliability in their laboratory environment.

Frequently Asked Questions

What class of safety do these cabinets provide?

The cabinets provide Class 100 cleanliness as per US Federal 209E, ensuring a sterile environment for critical laboratory work.

Is maintenance of these cabinets complicated?

Kalstein designs its equipment for easier maintenance. Regular filter checks and following specified cleaning protocols will ensure optimal performance.

Do these cabinets offer energy efficiency?

Yes, the cabinets are designed with energy efficiency in mind, having different maximum power consumption levels depending on the model, ranging from 400W to 700W.

Advantages and Disadvantages

The Kalstein Class II Biological Safety Cabinets offer numerous advantages including top-tier air cleanliness, energy efficiency, and a robust build boasting dimensions suitable for diverse laboratory setups. The low noise operation (≤65 dB(A)) also ensures a quiet working environment which is crucial for lengthy laboratory tasks.

One potential downside could be the initial investment cost, although this is offset by the long-term reliability and reduced maintenance needs. Moreover, the weight of the cabinets, although providing stability, may require thoughtful planning during installation.

Field Usage of the Product

In the field, these safety cabinets are utilized extensively in laboratories where the handling of dangerous microorganisms is routine. Microbiologists, virologists, and other laboratory professionals rely on these cabinets to maintain sterile environments and protect themselves and their samples from contamination.

These cabinets find applications in pharmaceutical companies, research institutions, and hospitals where biological safety is paramount. Their design focuses on preventing contamination through excellent ventilation and sterile surfaces.

Recommendations

To maximize the efficiency and lifespan of your Kalstein Class II Biological Safety Cabinet, routinely check the filters and ensure they are changed as per the manufacturer’s instructions. Cleaning should adhere to recommended procedures to maintain a sterile environment.

Position the cabinet in a location that minimizes airflow disruptions and ensure your laboratory staff is trained to use the equipment properly. Lastly, consider using Kalstein’s online portal for any enquiries or assistance regarding your cabinet model.

If you’re seeking a fusion of innovation and quality, you’ve come to the right place. At Kalstein, we offer you the luxury of exploring our exclusive catalog of laboratory equipment. We manufacture each piece with a level of excellence. Our intuitive and agile online purchasing channels are designed for your convenience, ensuring the most friendly prices. Don’t hesitate any more, we bring science to life, it’s time to be part of our community. https://kalstein.net/en/product/biological-safety-cabinet-100-exhaust-class-ii-a2-biosafety-cabinet-yr05704-yr05707/

We are here to help you

Send us a direct message

If you require a quote, you have a question about any product and / or want to clarify any doubt, contact us.

    PARIS

    Headquarter

    5 rue de Castiglione 75001 Paris – France.
    (+33) 01 70 39 26 50 (Line available in English and French languages)
    (+33) 7 60 83 92 83 (Line available in Spanish and French languages)
    Mail: [email protected]