Unveiling the Pulsed Xenon Lamp Spectrophotometer: Precision and Versatility

Product Description

The Pulsed Xenon Lamp Benchtop Spectrophotometer YR05475 is a cutting-edge instrument designed to deliver high precision and reliability in color measurement across diverse applications. Equipped with an advanced optical geometry for both reflectance and transmittance, this spectrophotometer operates with a d/8 (SCI & SCE; Include UV/Exclude UV) configuration for reflectance and a d/0 (SCI & SCE; Include UV/Exclude UV; Haze ASTM D1003) for transmittance. It adheres to international standards, including CIE No.15, GB/T 3978, ISO7724/1, ensuring accurate and consistent results.

The device boasts a 154mm integrating sphere, harnessing a light source that spans from 360nm to 780nm. Featuring a concave grating spectrophotometric mode and a 256 image element double array CMOS image sensor, it offers exceptional wavelength precision with a wavelength pitch of 10nm and a semiband width of 5nm. Designed to detect subtle differences in color, the spectrophotometer supports a wide range of color spaces, including CIE Lab, XYZ, s-RGB, among others.

The apparatus comes with a spectrum of apertures for varying sample sizes, including reflective apertures of Φ30mm/Φ25.4mm, Φ10mm/Φ8mm, and Φ6mm/Φ4mm. Its adaptability extends to the automatic identification of switch caliber, allowing for customized lens configurations. The measuring time is optimized to approximately 2.4 seconds, ensuring efficient workflow without compromising accuracy. With repeatability standards within 0.07% for spectral reflectance and spectral transmittance, alongside a chromaticity value repeatability within ΔE*ab 0.015, the YR05475 stands as a reliable choice for laboratory and industrial applications alike.

Model YR05475 (Pulsed xenon lamp)
Optical Geometry Reflectance: d/8 (SCI⪰ Include UV/Exclude UV)
Transmittance: d/0 (SCI⪰ Include UV/Exclude UV); Haze(ASTM D1003)
Conforms to CIE No.15, GB/T 3978, GB 2893, GB/T 18833, ISO7724/1, ASTM E1164, DIN5033 Teil7
Integrating Sphere Size Φ 154mm
Light Source Device 360nm-780nm Xenon Lamp, 400nm cut-off Xenon Lamp
Spectrophotometric Mode Concave Grating
Sensor 256 Image Element Double Array CMOS Image Sensor
Wavelength Range 360-780nm
Wavelength Pitch 10nm
Semiband Width 5nm
Reflectance Range 0~200%
Measuring Aperture Reflective: Φ30mm/Φ25.4mm, Φ10mm/Φ8mm, Φ6mm/Φ4mm;
Transmissive: Φ30mm/Φ25.4mm; Remark: 1. Automatic identification of switch caliber 2. Customized Configuration caliber and lens position
Specular Component Reflectance: SCI&SCE
Transmittance: SCI&SCE
Color Space CIE Lab, XYZ, Yxy, LCh, CIE LUV, Hunter LAB, Munsell, s-RGB, HunterLab, DIN, βxy
Color Difference Formula ΔE*ab, ΔE*uv, ΔE*94, ΔE*cmc(2:1), ΔE*cmc(1:1), ΔE*00v, ΔE(Hunter), DIN ΔE99
Colorimetric Index WI (ASTM E313, CIE/ISO, AATCC, Hunter), YI (ASTM D1925, ASTM 313), MI (Metamerism Index), Staining Fastness, Color Fastness, Color Strength, Opacity, Gardner Index, Pt-Co Index, 555 Index, Haze(ASTM D1003)
Observer Angle 2° & 10°
Illuminants D65, A, C, UV, D50, D55, D75, F1, F2, F3, F4, F5, F6, F7, F8, F9, F10, F11, F12
Displayed Data Spectrogram/Values, Chromaticity Values, Color Difference Values/Graph, Pass/Fail Result, Color Offset
Measuring Time About 2.4s (Measure SCI & SCE about 5s)
Repeatability Spectral reflectance: Φ25.4mm/SCI, Standard deviation within 0.07% (400 nm to 700 nm: within 0.06%)
Chromaticity value: Φ25.4mm/SCI, Standard deviation within ΔE*ab 0.015 (When a white calibration plate is measured 30 times at 5 second intervals after white calibration)
Spectral Transmittance: Φ25.4mm/SCI, Standard deviation within 0.07% (400 nm to 700 nm: within 0.07%)
Chromaticity value: Φ25.4mm/SCI, Standard deviation within ΔE*ab 0.018 (When a white calibration plate is measured 30 times at 5 second intervals after white calibration)
Inter-instrument Error Φ25.4mm/SCI, Within ΔE*ab 0.15
(Average for 12 BCRA Series II color tiles)
Size 370×300×200mm
Weight About 9.6kg
Power Supply DC 24V, 3A Power Adapter
Light Source Device Life 5 years, more than 3 million times measurements.
Display 7″ TFT Capacitive Screen-touch Display
Data Port USB & Print serial port
Data Storage Standard 2000 Pcs, Sample 20000 Pcs
Language Simplified Chinese, Traditional Chinese, English,
Working Environment Temperature: 0~40℃; Humidity: 0~85% (No Condensation)
Storage Environment Temperature: -20~50℃; Humidity: 0~85% (No Condensation)
Standard Accessory Power Adapter, User Guide, CD Disk(PC Software), USB cable, Standard Calibration Board, Black Calibration Cavity, Transmission black baffle, Sample holder, 25.4 caliber, 8 caliber, 4 caliber,
Optional Accessory Micro-printer, Transmissive Test Clamp Component,Micro Aperture(4mm) transmission test clamp component, Instrument inversion components
Notes: The specifications are subject to change without notice.

Market Price

In the ever-evolving landscape of laboratory equipment, the Pulsed Xenon Lamp Benchtop Spectrophotometer YR05475 positions itself as an affordable yet sophisticated option for precise color measurement needs. The market price generally falls within the range of 10,000 USD to 11,000 USD, reflecting its advanced capabilities and the superior quality it offers. This pricing balance makes it accessible for a wide range of industries seeking reliable color measurement without compromising on performance. When considering its feature-rich design and robust functionality, the investment in this spectrophotometer proves to be a wise choice for long-term value.

Frequently Asked Questions

What is the lifespan of the light source device?

The pulsed xenon lamp, which serves as the light source of the YR05475, has an impressive lifespan of 5 years, supporting over 3 million measurements, ensuring prolonged utility and reliability.

What kind of data can be displayed by this spectrophotometer?

The spectrophotometer is capable of displaying a vast array of data including spectrogram values, chromaticity values, color difference values/graphs, pass/fail results, and color offset. This diverse range allows users to gain a comprehensive understanding of the samples being analyzed.

Advantages and Disadvantages

The YR05475 spectrophotometer boasts several advantages, such as its high precision and repeatability, which are crucial for rigorous quality control processes. Its user-friendly interface via the 7″ TFT capacitive touch screen display streamlines the user experience, while the inclusion of multiple color spaces and observer angles enhances its versatility. Moreover, the device’s robust design and compliance with various international standards underscore its reliability and applicability across industries.

On the downside, the sophistication and number of features can present a learning curve for new users, requiring some time and training to fully leverage the spectrophotometer’s capabilities. Additionally, its high initial cost might be a consideration for smaller labs or enterprises with limited budgets. However, the long-term benefits in precision and operational efficiency often outweigh these initial considerations.

Product Use in the Field

In practical scenarios, the YR05475 spectrophotometer is indispensable for industries such as textiles, plastics, paint, and coatings, where precise color matching is critical. Its application extends to quality assurance departments that aim to maintain unwavering product consistency. The instrument’s capability to assess both reflectance and transmittance positions it as a versatile tool in research and development settings as well.

Real-world applications include the monitoring of color degradation over time in materials, optimizing formulations in product development, and ensuring compliance with international color standards. With its portable design and ease of use, this spectrophotometer is a valuable asset for on-site testing, providing quick, accurate results that inform crucial decision-making processes.

Recommendations

To maximize the utility and lifespan of the YR05475 spectrophotometer, it is essential to adhere to regular maintenance schedules, including timely calibration following the guidelines outlined in the user manual. Keeping the device and its components in optimal condition ensures consistent measurement accuracy.

Additionally, for new users, investing time in training and familiarization with the instrument’s functions can significantly improve the operation’s efficiency and effectiveness. Incorporating the use of the spectrophotometer into a structured workflow can leverage its full potential, aiding in achieving operational objectives with precision. Explore the full capabilities of this incredible device and enhance your laboratory’s performance by contacting us for a personalized quote or try our Kalstein Plus automatic quotation system to find the best deals available.

If you’re looking for a fusion of innovation and quality, you’ve come to the right place. At Kalstein we offer you the luxury of exploring our exclusive catalog of laboratory equipment. We manufacture each piece of equipment with a level of excellence. Our intuitive and agile online shopping channels are designed for your convenience, ensuring the most friendly prices. Don’t hesitate any longer; we give life to science, it’s time to become part of our community. https://kalstein.net/en/product/pulsed-xenon-lamp-benchtop-spectrophotometer-yr053405/.

We are here to help you

Send us a direct message

If you require a quote, you have a question about any product and / or want to clarify any doubt, contact us.

    PARIS

    Headquarter

    5 rue de Castiglione 75001 Paris – France.
    (+33) 01 70 39 26 50 (Line available in English and French languages)
    (+33) 7 60 83 92 83 (Line available in Spanish and French languages)
    Mail: [email protected]