Product Description
The biological safety cabinets YR05708, YR05709, and YR05710 are meticulously designed laboratory equipment engineered to offer high levels of cleanliness and safety. Adhering to Class 100@≥0.5μm standards as per US Federal 209E, these cabinets ensure an uncompromising level of contamination control. They maintain a colony count at ≤0.5pcs/dish·hour, utilizing a 90mm culture plate. With user-adjustable wind speeds, noise levels not exceeding 65dB(A), and robust voltage specifications of AC single-phase 220V/50Hz, these cabinets represent a blend of functionality and innovative design.
Each model offers varying specifications in terms of power consumption and weight, enhancing versatility in different laboratory settings. The control of air flow is facilitated by different average wind speeds that range across the cabinet’s interior, providing a controlled environment for sensitive laboratory work. For instance, inside the door the wind speed is maintained at 0.38±0.025m/s, in the middle it is 0.26±0.025m/s, and inside it’s 0.27±0.025m/s. The front suction wind speed is standardized at 0.55±0.025m/s.
Model | YR05708 | YR05709 | YR05710 | |
Cleanliness | Class 100@≥0.5μm (US Federal 209E) | |||
Number of colonies | ≤0.5pcs/dish·hour (Φ90mm culture plate) | |||
average wind speed | Inside the door | 0.38±0.025m/s | ||
middle | 0.26±0.025m/s | |||
Inside | 0.27±0.025m/s | |||
Front suction wind speed | 0.55m±0.025m/s | |||
noise | ≤65dB(A) | |||
Voltage | AC single-phase 220V/50Hz | |||
Maximum power consumption | 500W | 600W | 700W | |
Weight | 210KG | 250KG | 270KG | |
Working area W1×D1×H1 | 1040×650×620 | 1340×650×620 | 1640×650×620 | |
Overall dimension W×D×H | 1200×800×2270 | 1500×800×2270 | 1800×800×2270 | |
High efficiency filter specification and quantity | 980×490×50×1 | 1280×490×50×1 | 1580×490×50×1 | |
520×380×70×1 | 820×380×70×1 | 1120×380×70×1 | ||
Specification and quantity of fluorescent lamp/ultraviolet lamp | 20W×2/20W×1 | 30W×2/30W×1 | 30W×2/30W×1 |
Market Price
When considering the purchase of a biological safety cabinet of this caliber, it is essential to evaluate the pricing carefully. In the market, the price range for such sophisticated equipment typically falls between 6500 USD and 7700 USD. This range reflects the intricate design, advanced features, and high performance that this type of laboratory equipment offers. It’s a worthy investment for any lab prioritizing safety and reliability in their operations.
Frequently Asked Questions
What is the primary function of a biological safety cabinet?
A biological safety cabinet is designed to provide a contaminant-free environment for working with sensitive microorganisms. It ensures the protection of the user, the environment, and the product itself.
How often should the filters in a biological safety cabinet be replaced?
The lifespan of cabinet filters varies based on usage and environment. Generally, it is recommended to replace filters every 12-18 months or sooner if performance begins to decline.
Advantages and Disadvantages
Advantages: The biological safety cabinets offer unparalleled protection, enhanced by their high efficiency filtration system and stable airflow. They provide a controlled working space, limiting the risk of contamination. The ergonomic design and quiet operation (≤65dB(A)) also improve user comfort.
Disadvantages: The primary disadvantage could be the initial investment cost, which may be substantial for some laboratories. Moreover, regular maintenance and filter replacements are necessary to maintain optimal performance, which could add to long-term costs.
Use of Product in the Field
These biological safety cabinets are predominantly used in microbiological and biomedical research laboratories. They are essential for tasks involving pathogenic agents or requiring sterility, such as tissue culture work, genetic research, and pharmaceutical testing. The precision in airflow management makes them invaluable where contamination prevention is critical.
In clinical settings, biological safety cabinets play a vital role in ensuring that both samples and personnel are protected from potential biohazards. Their reliability in maintaining a sterile environment helps streamline processes and boost productivity.
Recommendations
To ensure the biological safety cabinets deliver optimal performance and longevity, it is crucial to conduct regular maintenance checks. Ensure the filters are functioning correctly and replace them as needed. Also, training users on the proper operation of the cabinets will minimize the risk of misuse and potential contamination.
Incorporating regular calibration of the airflow velocities and monitoring noise levels can ensure the continued efficacy of the equipment. Taking these precautionary steps will help in maintaining a sterile environment crucial for successful laboratory operations.
If you are searching for a fusion of innovation and quality, you have come to the right place. At Kalstein, we offer you the luxury of exploring our exclusive catalog of laboratory equipment. We manufacture each piece with a high level of excellence. Our intuitive and agile online purchase channels are designed for your convenience, ensuring the most affordable prices. Don’t hesitate any longer; we bring science to life, and it’s time to be part of our community. https://kalstein.net/en/product/biological-safety-cabinet-yr05708-yr05710/.